Liquiline M CM42
Two-wire transmitter for pH measurement with digital IS-FET sensors
Part 1: Commissioning
BA381C/07/en/PODValid as of:
Software version 10.04.00
Overview of documentation supplied
Operating Instructions
The Operating Instructions are split into two parts:•Commissioning
–Description of your device
–All steps that you have to carry out once only during initial commissioning•Operation
–Software description–Troubleshooting
–All steps that you have to carry out during the routine operation of your device, depending on your measuring task
!
Note!
This part of the Operating Instructions is the documentation on commissioning.
CD-ROM
Contents:
•Fundamentals of pH and conductivity measurement•Communication
•Overview of analytical measuring technology•Accessories•Safety
•Ordering information
In addition
•1 manufacturer's certificate•Adhesive labels:
–Replacement nameplate (landscape format)–Wiring adhesive label
Liquiline M CM42
Table of contents
1
Safety instructions . . . . . . . . . . 4
1.1 Designated use . . . . . . . . . . . . . . . . . . . . . . 41.2 Incoming acceptance, transport, storage . . . 41.3 Installation, commissioning and operation . 41.4 Operational safety . . . . . . . . . . . . . . . . . . . 51.5 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51.6
Notes on safety conventions and icons . . . . 5
2
Identification . . . . . . . . . . . . . . 6
2.1 Device designation . . . . . . . . . . . . . . . . . . . 62.2 Scope of delivery . . . . . . . . . . . . . . . . . . . . 82.3 Reordering documentation . . . . . . . . . . . . . 82.4
Certificates and approvals . . . . . . . . . . . . . . 9
3 Device description . . . . . . . . . 103.1 Housing closed . . . . . . . . . . . . . . . . . . . . . 103.2 Housing open . . . . . . . . . . . . . . . . . . . . . . 10
4
Installation. . . . . . . . . . . . . . . 11
4.1 Measuring system . . . . . . . . . . . . . . . . . . 114.2 Installation conditions . . . . . . . . . . . . . . . 124.3 Installation instructions . . . . . . . . . . . . . . 144.4
Post-installation check . . . . . . . . . . . . . . . 16
5 Wiring . . . . . . . . . . . . . . . . . . 17
5.1 Electrical connection . . . . . . . . . . . . . . . . 175.2 Post-connection check . . . . . . . . . . . . . . . 22
6
Commissioning. . . . . . . . . . . . 23
6.1 Function check . . . . . . . . . . . . . . . . . . . . 236.2 Configuring the device address . . . . . . . . . 236.3 Quick Setup (local operation) . . . . . . . . . . 236.4 User administration . . . . . . . . . . . . . . . . . 266.5
Calibration . . . . . . . . . . . . . . . . . . . . . . . . 27
7 Maintenance. . . . . . . . . . . . . . 298
Troubleshooting - hardware. . 30
8.1 Replacing modules . . . . . . . . . . . . . . . . . . 308.2 Spare parts . . . . . . . . . . . . . . . . . . . . . . . . 318.3 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . 338.4
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Endress+Hauser
9
Accessories. . . . . . . . . . . . . . .34
9.1 Installation kits . . . . . . . . . . . . . . . . . . . . . 349.2 Weather protection . . . . . . . . . . . . . . . . . . 349.3 Communication . . . . . . . . . . . . . . . . . . . . 349.4
Data management . . . . . . . . . . . . . . . . . . . 34
10 Technical data. . . . . . . . . . . . .35
10.1 Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3510.2 Output . . . . . . . . . . . . . . . . . . . . . . . . . . . 3510.3 Electrical connection . . . . . . . . . . . . . . . . . 3610.4 Performance characteristics . . . . . . . . . . . . 3610.5 Environment . . . . . . . . . . . . . . . . . . . . . . 3610.6
Mechanical construction . . . . . . . . . . . . . . 37
Index. . . . . . . . . . . . . . . . . . . .38
Safety instructionsLiquiline M CM42
1
1.1
Safety instructions
Designated use
Liquiline CM42 is a two-wire transmitter for liquid analysis in all areas of process engineering:•Chemical processes•Pharmaceutical industry•Food technology
If the device is used for any purpose other than that described, this poses a threat to the safety of people and the entire measuring system and is thus not permitted.
The manufacturer does not accept liability for damage caused by improper or non-designated use.
1.2Incoming acceptance, transport, storage
•Make sure the packaging is not damaged!
Inform your supplier of any damage to the packaging.
Please keep the damaged packaging until any issues have been resolved.•Make sure the contents are not damaged!
Inform your supplier of any damage to the contents.
Please keep the damaged goods until any issues have been resolved.
•Check the scope of delivery against the delivery papers and your order to ensure it is complete and nothing is missing.
•Pack the product in such a way as to protect it reliably against impact and moisture for storage and transportation. Optimum protection is provided by the original packaging materials. In addition, the permitted ambient conditions must be observed (see Technical data).•If you have any queries, please contact your supplier or local sales center.
1.3Installation, commissioning and operation
Observe the following points:
•Installation, commissioning, operation and maintenance of the measuring system must only be carried out by trained technical personnel.
The technical personnel must be authorized to perform the tasks by the owner-operator.•The electrical connection may only be established by an electrical technician.
•The technical personnel must have read and understood these Operating Instructions and must follow the instructions they contain.
•Prior to commissioning the entire measuring point, check that all connections are correct. Make sure that electric cables and hose connections are not damaged.
•Do not commission damaged products. Protect them against unintentional startup. Label and identify the damaged product as defective.
•Faults at the measuring point may only be rectified by authorized and properly trained personnel.
4Endress+Hauser
Liquiline M CM42Safety instructions
•If the faults cannot be eliminated, take the products out of service and protect them against unintentional startup.
•Repairs not described in these Operating Instructions may only be carried out directly at the manufacturer's or by the Service Organization.
1.4Operational safety
The transmitter is designed to meet state-of-the-art safety requirements, has been tested and left the factory in a condition in which it is safe to operate.
The applicable regulations and European standards have been taken into account.As the user, you are responsible for observing the following safety regulations:•Installation guidelines
•Local standards and regulations
1.5Return
If returning the transmitter, please send it cleaned to your sales center.Use the original packaging when returning the device.
1.6Notes on safety conventions and icons
1.6.1Warnings
Warning!
This symbol alerts you to dangers.
If the warning is not observed, this can result in serious injury to people or damage to property.
Caution!
This symbol alerts you to possible malfunctions resulting from incorrect operation. If not observed, this could result in damage to property.Note!
This symbol indicates important items of information.
#\"!
1.6.2Document symbols
È ä 1This symbol stands for a cross-reference to a certain page (e.g. Page 1).È å 2This symbol stands for a cross-reference to a certain graphic (e.g. Fig. 2).
Endress+Hauser5
IdentificationLiquiline M CM42
2
2.1
Identification
Device designation
2.1.1Nameplate
Nameplates can be found:
•On the outside of the housing (lasered)
•On the packaging (adhesive label, portrait format)
•On the adhesive label sheet provided (landscape format, for your own use)Compare the data on the nameplate with your order.
The nameplate provides you with the following information on your device:1.Order code2.Serial number3.Device version4.Measuring range(s)5.Input, output values
6.Approvals (depending on the version ordered)7.Protection class
8.Ambient temperature range9.Approval symbols
a0003281Fig. 1: Nameplate information
10.Safety notices, warnings
2.1.2Serial number and order code
The order code and serial number of your device can be found in the following locations:•On the nameplate
•On the front page of the Operating Instructions•In the delivery papers
!
Note!
•The documentation only contains information that was valid at the time of delivery.•If you have any queries, please contact your local sales center.
6Endress+Hauser
Liquiline M CM42Identification
2.1.3Product structure
Sensor inputANomoduleCConductivity, conductive measurement, analog sensorIConductivity, inductive measurement, analog sensorKDigital sensor: conductivity, conductive measurementMDigital sensor: pH/ORP measurement, glass sensorNDigital sensor: pH measurement, ISFET sensorODigital sensor: amperometric oxygen measurementPpH (glass/ISFET) or ORP, analog sensor
ApprovalANon-hazardous area, CSA GPGAtex II (1)2 G EEx ia/ib IIC T6 / II 3 DHAtex II 3G EEx nA II T6NNEPSIExnAIIT6PFM IS NI Cl. I, Div. 1&2, Groups A-D
SCSA IS NI Cl. I, II, III, Div. 1&2, Groups A-G
Certificate ANot selectedBTest certificate to EN10204, 3.1CTest certificate to EN10204, 3.1, factory calibration certificate
Output
® 01 current output 4 to 20 mA, HART
12 current outputs 4 to 20 mA, 1x HART®2PROFIBUSPA3FOUNDATION Fieldbus
Housing
0Plastic housing1Stainless steel housing 1.4404 (AISI 316L)
Cable entry0M20 x 1.51NPT½\"2G½
Software EAStandard versionEBAdvanced versionECAdvanced Plus version
Device languageDDe/En(Docs,German)EEn/De (Docs, English)FFr/En(Docs,French)SEs/En (Docs, Spanish) ZZh/En (Docs, Chinese)
Documentation0Installation + operation
Additional fittings0Basic version
Complete order code
CM42-!
Note!
Please ask your sales office for other device versions, e.g. other software or documentation languages.
Endress+Hauser7
IdentificationLiquiline M CM42
2.2
••••••
Scope of delivery
1 transmitter in the version ordered
1 wall securing unit incl. 4 flat head screws
1 sheet of adhesive labels (nameplates, terminal connection diagrams)1 Operating Instructions Part 1, BA381C \"Commissioning\"1 Operating Instructions Part 2, BA382C \"Operation\"1 CD-ROM with additional documentation
If you have any queries, please contact your supplier or local sales center.
2.3Reordering documentation
What to order?
•Reordering lost documentation or documentation of which several copies are required:You must quote the serial number of your device.
You receive the same documentation as received with the delivery.
•First time ordering appropriate documentation for devices which you have assembled yourself by installing a module:
You have to indicate the order code the device corresponds to after installing the module. Specify a serial number also.
You receive the documentation to suit your newly modified device.How to order?
•By Internet (www.endress.com/liquiline-documentation →å2):You receive PDF files of both parts of the Operating Instructions.•Through your sales center:
Indicate you want the documentation as hard copy.
a0006062
Fig. 2: www.endress.com/liquiline-documentation
8Endress+Hauser
Liquiline M CM42Identification
2.4Certificates and approvals
With this declaration, the manufacturer guarantees that the product conforms to the regulations of European Directive 89/336/EEC. This is proven by observing the standards listed in the declaration of conformity.
Endress+Hauser9
Device descriptionLiquiline M CM42
3
3.1
Device description
Housing closed
1345-8
Wall securing plateAlarm LEDDisplayNavigatorSoft keys,
assignment depends on menu
122
348Fig. 3: CM42 with closed display cover
765a00010673.2Housing open
a0001069Fig. 4: 123With open display cover (without wiring)DAT memory moduleCPU moduleSensor module
456DIN rail
Threaded joints
CDI socket (service interface)
10Endress+Hauser
Liquiline M CM42Installation
4
4.1
Installation
Measuring system
A complete measuring system consists of:•Liquiline M CM42
•An assembly, e.g. fixed installation assembly CPA442•A digital sensor, e.g. Tophit CPS471-D•A measuring cable CYK10
a0001399Fig. 5: Example of a measuring system1234
Liquiline M CM42
Measuring cable CYK10Digital sensor
Fixed installation assembly CPA442
Endress+Hauser11
InstallationLiquiline M CM42
4.2Installation conditions
4.2.1Dimensions
a0001074Fig. 6: Dimensions
12Endress+Hauser
Liquiline M CM42Installation
4.2.2Mounting plate
a0001075Fig. 7: Mounting plate
4.2.3Weather protection cover
!
Note!
When installing outside, always use the weather protection cover (see 'Accessories', CD-ROM) to protect the transmitter against rain and snow.
107(4.21)90°253(9.96)224(8.82)98(3.86)151(5.94)Fig. 8: Weather protection cover
98(3.86)mm(inch)a0001671Endress+Hauser
230(9.01)100°13
InstallationLiquiline M CM42
4.3Installation instructions
4.3.1Wall or field mounting
There are many ways of mounting the device:•Wall mounting
•Mounting on a vertical pipe or post (round or square)
•Mounting on railing or on a horizontal pipe (round or square)
!
Note!
You require the post mount to mount on a pipe or post. This is an accessory and is not included in the scope of delivery (see \"Accessories\").
121244Fig. 9: Mounting on a vertical pipe/post12
CM42 (as delivered, incl. a mounting plate)
Securing screws (part of the post mount, see Accessories)
3a00030923a0002162Fig. 10: Side view34
Mounting plate (post mount)Pipe or post (round/square)
14Endress+Hauser
Liquiline M CM42Installation
121244Fig. 11: Mounting on a horizontal pipe/railing12
CM42 (as delivered)
Securing screws (post mount)
3a00030933a0002164Fig. 12: Side view34
Mounting plate (post mount)Pipe or railing
a0002166Fig. 13: Wall mounting (optional weather protection cover)
4.3.2Panel mounting
!
Note!
•For panel mounting, you need the installation kit consisting of tensioning screws and a front seal. This is an accessory (see \"Accessories\") and is not included in the scope of delivery.•If installing the units above one another, you must observe a minimum distance for the cable glands of the upper device.
•If installing the units beside one another, you must observe a minimum distance for opening the front of the housing.
•If arranging in a square, you must take into account the lengths of the mounting plates at the rear of the device or the cable glands for minimum spacing distance.
Endress+Hauser15
InstallationLiquiline M CM42
a0002402Fig. 14: Panel mounting: left side view, right front view
4.4Post-installation check
•After installation, check the transmitter for damage.
•Check whether the transmitter is protected against moisture and direct sunlight.
16Endress+Hauser
Liquiline M CM42Wiring
5
5.1
Wiring
Electrical connection
#
Warning!
•The electrical connection may only be established by an electrical technician.
•The electrical technician must have read and understood these Operating Instructions and must follow the instructions they contain.
•Prior to beginning any wiring work, make sure voltage is not applied to any of the cables.5.1.1Opening the housing1.
Loosen the four screws on the front with a Phillips screwdriver:
2.Open the housing:
\"
Caution!
Do not use any sharp or pointed objects, such as a screwdriver, knife etc., to open the housing. Otherwise, you might damage the housing seal, scratch the housing etc.
Endress+Hauser17
WiringLiquiline M CM42
5.1.2Housing grounding
\"
Caution!
You must connect the \"fixing plate\" to the foundation ground with a separate ≥2.5mm2 (i14 AWG) functional ground.
a0003617Fig. 15: Grounding the housing
12
Fixing plate
≥2.5mm2 (i14 AWG) functional ground
5.1.3Cable grounding in housing
!
Note!
•If possible, only use terminated genuine cables.
•Ground all the cables in the housing of the transmitter as shown in the following diagrams (sample cable, does not necessarily correspond to the original cable).123a0002679a0002680a0002681Fig. 16: Terminated cable (example)123
Outer shield (exposed)Cable cores with ferrulesCable sheath (insulation)
Fig. 17: Inserting the cable4
Grounding clip
Fig. 18: Grounding the cable
(functional ground)
18Endress+Hauser
Liquiline M CM42Wiring
5.1.4Cable terminals
1.Insert a suitable screwdriver into the opening
of the terminal spring (square opening) until the stop.
a0003602Fig. 19: Opening the terminal
2.Insert the terminated cable end into the
terminal opening (round opening).
a0003603Fig. 20: Inserting the cable
3.Remove the screwdriver. Make sure the cable
is securely positioned in the terminal.
a0003604Fig. 21: Releasing the terminal spring
5.1.5Signal output / power supplyPermitted supply voltage
Current output 4 to 20 mA without HART®:Current output 4 to 20 mA with HART®:
1)
12.5 to 30 V DC12.5 to 30 V DC1)
Permitted residual ripple to IEC 60381-1:1985 Par. 3.3: < 100 mVpp in frequency range 50 Hz to 3 kHz
Connection in transmitter
!
Note!
Connect the transmitter with two-wire cabling shielded on both sides.
Endress+Hauser19
WiringLiquiline M CM42
a0002365Fig. 22: View in device (CPU module)
a0003100Fig. 23: Wiring diagram
Connection via HART interface
12RS232³250W+-1331343a0004595Fig. 24: Connection: HART without PLC123
Power unit
HART modem Commubox FXA191HART handheld terminal
20Endress+Hauser
Liquiline M CM42Wiring
1+-1331342RS2323a0004597Fig. 25: Connection: HART with PLC123
Additional switching units or PLC with passive inputHART modem Commubox FXA191HART handheld terminal
5.1.6Sensor connection
Explanation of abbreviations in the following diagrams:AbbreviationU+U–
Com ACom B
Meaning
Signals of the digital sensor
!
Note!
•Cable colors indicated as per IEC 757 (see CD–ROM).
•You must connect shielded connections and terminals with functional ground ()) (there is no protective ground (*) for plastic housings).
Endress+Hauser21
WiringLiquiline M CM42
a0006128a0001078Fig. 26: View in device (sensor module)Fig. 27: Wiring diagram
5.2
1.2.3.4.5.6.7.
Post-connection check
Are the transmitter and cables free from damage on the outside?Are the mounted cables strain relieved?Are the cables run without loops and cross-overs?
Are the signal lines correctly connected in accordance with the wiring diagram?Have you positioned unused connection wires on the \"fixing plate\"?Are all the connection wires securely positioned in the cable terminals?Are all the cable entries installed, tightened and sealed?
Device state and specificationsElectrical connection
22Endress+Hauser
Liquiline M CM42Commissioning
#
Warning!
Only put the transmitter into operation if you can answer yes to all these questions. Otherwise, safe operation cannot be guaranteed. In such instances, the manufacturer does not accept any liability for the device.
6
6.1
Commissioning
Function check
#
Warning!
•Check that all connections have been established correctly.
•Make sure that the supply voltage matches the voltage indicated on the nameplate!
6.2Configuring the device address
Factory setting: 0.
You can change the device address to integrate several HART devices in a network (Multidrop mode).
Use the handheld terminal (DXR375) to change the device address.
6.3
1.2.3.
Quick Setup (local operation)
Connect the supply voltage.
Wait for the initialization to be complete.
Completely work through the \"Quick Setup\" submenu in \"PARAM\".
6.3.1Menu structure
Function name of display (local operation)PARAMQuick Setup
LanguageDate formatSet dateTime formatSet timeTemperature unitBuffer manufacturerTemp. compensation
Endress+Hauser23
CommissioningLiquiline M CM42
Function name of display (local operation)Medium temperature(Temp. compensation=\"Man. comp.\")Current output 1
Output sourceLow value (4mA)Upper value (20mA)Current output 2
Output sourceLow value (4mA)Upper value (20mA)
6.3.2Configuration options
The abbreviation \"AC\" is used in the table header. AC stands for \"Access Controls\write authorization.
The table only indicates the rights of the \"Maintenance\" user group as the \"Specialist\" group has administrator status and thus always has full rights (--> \"Commissioning/User administration\").•R = Read only, i.e. the maintenance engineer can only read the data
•R/W = Read/Write, i.e. as a maintenance engineer, you can also change values.
!
Note!
The access controls do not take effect until password protection has been activated!
FunctionLanguage
Options
Options•English
•Language orderedFactory settingLanguage ordered
Date format
Options
•DD.MM.YYYY•MM.DD.YYYYFactory settingDD.MM.YYYY
Set dateTime format
Depends on the formatDD.MM.YYYYOptions
•hhmmss (24 h)•hhmmss (am / pm)Factory settinghhmmss (24 h)
Set time
Depends on the formathh:mm:ss
R/WR/WR/W
24-hour display or 12-hour displayEditing mode:
hh (hour): 0 to 23 / 0 am to 12 pmmm (minutes): 0 to 59ss (seconds): 0 to 59
R/WACR/W
Info
Note! !\"Language ordered\" refers to the language that you selected for your device using the order code (\"device language\").
If you select the other language in question, all the other settings remain intact.Editing mode:DD (day): 1 to 31MM (month): 1 to 12
YYYY (year): 2005 to 2100
24Endress+Hauser
Liquiline M CM42Commissioning
FunctionTemperature unit
OptionsOptions• °C• °F
Factory setting °C
ACR/W
Info
Buffer manufacturer
Options
•E+H(NIST)•Ingold/Mettler•DIN 19267•DIN 19266•Merck/Riedel Factory settingE+H (NIST)
R/W
Temp. compensation
Options•Off
•Autocomp.(ATC)•Man.comp.Factory settingAuto comp. (ATC)
R/W
Medium temperature
–50 to + 250 °C
(–58 to + 482 °F)Factory setting25 °C (77 °F)
R/W
Only for temp. compensation = \"Man. comp.\"
Current output 1Output source
Options
•Mainvalue
Factory setting Main value
–2 to
(20 mA value) - 1 pHFactory setting2 pH
Upper value (20mA)
(4 mA value) + 1 pH to16 pHFactory setting12 pH
Current output 2Output source
Options
•Mainvalue•TemperatureFactory settingTemperature
RR/WR
Main value
Low value (4mA)R/W
Specify the measured value that should correspond to the 4mA or 20mA value.The smallest possible spread between4mA and 20mA value is:1 pH
Temperature
Endress+Hauser25
CommissioningLiquiline M CM42
Function
Low value (4mA)
Options
–50 to
(20 mA value) - 5 °CFactory setting-20 °C
ACR/W
Info
The smallest possible spread between 4mA and 20mA value is 0.1 °C.
Upper value (20mA)
(4 mA value) + 5 °C to250 °CFactory setting+100 °C
R/W
6.4User administration
To protect the device against unwanted or unauthorized access, you can assign an access code to the menu.
!
Note!
The factory setting for the access code is an empty field (= no entry). The prompt to enter a password is deactivated. Setting is possible for everyone.Assigning/changing access codes:1.2.
Go to PARAM menu, \"General settings\"/\"User administration\".Select the type of protection:
–No password protection (= factory setting).–Password protection per code entry.
If you select \"Enter code\roles.
3.
6.4.1User roles and access authorization
If the password protection has been defined and activated, you have to log in with the appropriate user role before you can make changes to the settings in the menu.
26Endress+Hauser
Liquiline M CM42Commissioning
Here, you will be prompted by the device to enter an access code as soon as you want to configure a value:
User roles and access authorization (factory setting):
Maintenance
•Read and, sometimes, write access in the PARAM and DIAG menus
•Read and write access in the calibration menu
Specialist
Admin status:
•Read and write access in all menus
•Change the access authorization for other user groups
!
Note!
If you are logged in as the \"Specialist\" (= Administrator), you can change the access authorization of all other user groups:1.2.
Select a software function for which you want to assign authorization to, or take authorization from, another user group.
Press and hold down the navigator button for at least 3 seconds.
The display shows all the user groups and the current access authorization to the function selected.
Use the navigator to change the authorization in question and confirm with \"OK\".Make a note of the changes in your operating logbook or similar.
3.4.
6.5Calibration
Calibration is necessary:•After periods of standstill
•At sensible, process-dependent intervals1)
1) Depending on the process conditions, the intervals can range from several times daily to once quarterly.
Endress+Hauser27
CommissioningLiquiline M CM42
6.5.1Types of calibration
•Two-point calibration
–With automatic buffer recognition:
You select the buffer type (e.g. DIN 19266). The device automatically recognizes the buffer during calibration.–With fixed buffer:
You select two buffer solutions from the menu.•Single-point calibration –You enter an offset.•Numeric input:
–You enter the slope, zeropoint and temperature.•Grab sample calibration
–You take a sample of the medium. The device determines a measured value while you analyze the sample in the laboratory.
Then you enter the laboratory measured value. The device calibrates on the basis of the laboratory value.6.5.2Calibrate
Proceed as follows to calibrate the sensor:1.2.3.4.5.6.
Remove the sensor from the process.Clean the sensor.
Press the soft key for \"CAL\".Follow the instructions in the menu.
Finish calibrating by switching back to the measuring mode.Install the sensor back into the process.
Your measuring point is now ready for operation.
!
Note!
•If calibration is aborted using ESC, or if the calibration is faulty, the system continues to use the original calibration data. A calibration error is shown as plain text on the display. •Any offset set is automatically deleted after accepting the calibration.
28Endress+Hauser
Liquiline M CM427Maintenance
Clean the front of the housing with usual commercial cleaning agents.The front is resistant to the following in accordance with DIN 42 115:•Alcohol (briefly)
•Diluted acids (max. 2% HCl)•Diluted alkalis (max. 3% NaOH)•Soap-based household cleaner
\"
Caution!
Never use any of the following for cleaning purposes:•Concentrated mineral acids or alkalis•Benzyl alcohol•Methylene chloride•High-pressure steam
Endress+HauserMaintenance
29
Troubleshooting - hardwareLiquiline M CM42
8
8.1
Troubleshooting - hardware
Replacing modules
Exchange modules when replacing defective modules or in the event of changes to the configuration of your device.
112a0004575a0004577Fig. 28: Removing module1
Removal aids
Fig. 29: Housing guides12
Guides on housing wallGuidesinDINrail 1a0004578a0004579Fig. 30: Module, left side (cable connections = front)1
Guides (e.g. CPU module to fit into the guides in
housing)
Fig. 31: Module, right side (cable connection = front)12
Guides (e.g. input module to fit into the right-hand guides of the CPU module)Removal aids
1.2.3.4.5.6.7.
Disconnect Liquiline from the power supply and open the housing.Remove the cable connections from the module you want to replace.Pull out the two removal aids on the module until the stop (→å28).The module can now be easily removed from the DIN rail.
Slide the new module into the guides (→å29, →å30, →å31).
Push the two removal aids on the module in the direction of the DIN rail until the stop. This locks the module onto the DIN rail.
Connect the cables in accordance with the wiring diagram (see \"Wiring\").
Close the housing, connect the sensor and check that the entire measuring system is working correctly.
30Endress+Hauser
Liquiline M CM42Troubleshooting - hardware
8.2Spare parts
a0004474Fig. 32: Exploded view
Please refer to the following table for item names and order numbers for spare parts.
Endress+Hauser31
Troubleshooting - hardwareLiquiline M CM42
Item1
Kit CM42
Housing base, plastic, M20 1)•Base
•Plate M20 (item 15)•Fixing plate cpl. SS
•3 threaded joints M20x1.5 (item 14)
CPU module, curr. outp. + HART, non-Ex
•FMIH1 insertion module (FC2W1+FBIH1), non-Ex•Terminal strip (item 3)
Terminal set, CPU module, curr. outp. + HARTRibbon cable for plastic housing
•1 cable, conf. 40xAWG28 EX, 90 mm (3.5\") crimpHinge for plastic housing•Hinge cpl. CM42 KS
•2 special screws K40x10 Torx self-form.DATchipwith software
•FMDC1 FTC electronics BG EX
Top housing section, plastic, with display
•Display, keys, navigator, screws and cover plate•Hinge(item5)
•Ribbon cable (item 4)
Inp. module conductivity, conductive measurement, non-Ex•FSLC1, insertion module, cond. conductivity, non-Ex
•Terminal strip (item 9)
Order number71001361
251518002
345
515174815151750351517500
67
5151750951517462
851518005
8
Inp. module conductivity, inductive measurement, non-Ex •FSLI1, insertion module, ind. conductivity, non-Ex
•Terminal strip (item 9)
Inp. module terminal set, conductivity, conductive measurementInp. module terminal set, conductivity, inductive measurementInp. module digital/Memosens, non-Ex
•FSDG1, insertion module, 1 channel, 1 SWU, non-Ex•Terminal strip (item 11)
Inp. module terminal set, digital/MemosensInp. module pH/ORP/temperature, non-Ex•FSPH1, insertion module, pH/mV, non-Ex•Terminal strip (item 13)
Inp. module terminal set, pH/ORP/temperatureThreaded joints M20•1 set of threaded joints•Thread base plate
Terminal set, 5 pcs. each for all modules, 7 types
51518006
9910
515174895151749051518007
1112
5151749151518004
1314
5151748751517504
No Fig.51517498
32Endress+Hauser
Liquiline M CM42Troubleshooting - hardware
ItemKit CM42
Order numberNo Fig.
10 set cable clamps+screws
51517499
•10 EMC cable clamps D6 (0.24\")•10 EMC cable clamps D4 (0.16\")•10 cheese head screws M4x8 A2No Fig.Connection jack for external Historom/CDI
51517507
1)
When ordering, you must specify the serial number of the device for which you are ordering the spare part.
8.3Return
If returning the transmitter, please send it cleaned to your sales center.Use the original packaging when returning the device.
8.4Disposal
This product contains electronic components. For this reason, it must be disposed of as electronic waste.
Please observe local regulations.
Endress+Hauser33
AccessoriesLiquiline M CM42
9Accessories
!
Note!
The CD-ROM supplied provides you with all the relevant information on an extensive range of accessories, including descriptions, graphics and ordering information.For example, it contains information on:•Sensors•Assemblies
•Quality buffers from Endress+Hauser
9.1Installation kits
Post mount
•1 Mounting plate•2 Securing screws•Order No. 51518263
Panel installation kit for panel cutout 138x138 mm (5.43x5.43 inch)•1 Panel installation seal
•2 Tensioning screws M6x150 mm•4 Hexagonal nuts M6, DIN934 A2•4 Spring washers, A2 DIN127, form B6•4 Washers A6.4, DIN125 A2•Order No. 51518173
9.2Weather protection
Weather protection cover •Order No. 51517382
9.3Communication
HART handheld terminal DXR375
•For communicating with a HART-compatible device via a 4 to 20 mA line•Order No. DXR 375
HART modem Commubox FXA291
•Interface module between HART and serial PC interface•Technical Information TI405C/07•Order No. 51516983
9.4Data management
Memobase CYZ41D
•Software for central data and sensor management•for Windows 2000 or XP
•Ordering as per product structure
34Endress+Hauser
Liquiline M CM42Technical data
10Technical data
10.1Input
10.1.1Measured variables•pH value•Temperature
10.1.2Measuring rangepH valueTemperature
0 to 14
–25 to +150 °C (–10 to 300 °F)
10.1.3Cable specificationWith Memosens (CYK10)
Max. cable length 100 m (330 ft)
10.2Output
10.2.1Output signal
2x 4 to 20 mA, potentially isolated against sensor circuit0.8 to 1.2 mA peak to peak2)10.2.2Signal on alarm≥21.5 mA10.2.3Load
Max. load with supply voltage of 24V: 500 ΩMax. load with supply voltage of 30 V: 750 Ω10.2.4Output distributionpH valueTemperature
Configurable, ΔpH > 0.5Adjustable, at least 2 °C (2 °F)
2) Only for HART
Endress+Hauser35
Technical dataLiquiline M CM42
10.3Electrical connection
10.3.1Supply voltage and signal voltageCurrent output 4 to 20 mA without HART®:Current output 4 to 20 mA with HART®:
1)
12.5 to 30 V DC12.5 to 30 V DC1)
Permitted residual ripple to IEC 60381-1:1985 Par. 3.3: < 100 mVpp in frequency range 50 Hz to 3 kHz
10.3.2Cable cross-section
Max. cable cross-section: 2.5 mm2 (i14 AWG), GND 4 mm2 (i12 AWG)
10.4Performance characteristics
10.4.1Measured value resolutionpH
Leakage current, ISFET sensorsTemperature
0.01 pH100 nA0.1 °C (0.1 °F)
10.4.2Maximum measured error3)Current outputs, additionalpH
Leakage current, ISFET sensorsTemperature
20 μA0.02 pH100 nA1 K
10.5Environment
10.5.1Ambient temperature range–30 to 70 °C (–20 to 160 °F)
3) In accordance with IEC 746-1 for nominal operating conditions
36Endress+Hauser
Liquiline M CM42Technical data
10.5.2Ambient temperature limits–30 to +80 °C (–20 to 175 °F)10.5.3Storage temperature–40 to 80 °C (–40 to 175 °F)
10.5.4Electromagnetic compatibility
Interference emission and interference immunity to EN 61326: 200410.5.5Degree of protectionIP67 (iNEMA 4X)10.5.6Relative humidity10 to 95%, not condensing
10.6Mechanical construction
10.6.1Weight1.5 kg (3.3 lbs)10.6.2MaterialHousing
Housing seals
PolycarbonateEPDM
Endress+Hauser37
Liquiline M CM42
Index
A
Access authorization
Change. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26User roles. . . . . . . . . . . . . . . . . . . . . . . . . . . . 26Access code
Factory setting . . . . . . . . . . . . . . . . . . . . . . . . 26Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34Communication . . . . . . . . . . . . . . . . . . . . . . . 34Installation kits. . . . . . . . . . . . . . . . . . . . . . . . 34Ambient temperature limits. . . . . . . . . . . . . . . . . 37Ambient temperature range. . . . . . . . . . . . . . . . . 36Approvals. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
B
Bus address. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
C
Cable cross-section. . . . . . . . . . . . . . . . . . . . . . . 36Cable grounding. . . . . . . . . . . . . . . . . . . . . . . . . 18Cable specification . . . . . . . . . . . . . . . . . . . . . . . 35Cable terminals. . . . . . . . . . . . . . . . . . . . . . . . . . 19Calibrate. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27Check
Electrical connection. . . . . . . . . . . . . . . . . . . . 22Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23Post-installation . . . . . . . . . . . . . . . . . . . . . . . 16Commissioning. . . . . . . . . . . . . . . . . . . . . . . . 4, 23Communication . . . . . . . . . . . . . . . . . . . . . . . . . 34Configuring
Device address . . . . . . . . . . . . . . . . . . . . . . . . 23Connecting ground. . . . . . . . . . . . . . . . . . . . . . . 18Connection
Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22Connection data. . . . . . . . . . . . . . . . . . . . . . . . . 36
D
Degree of protection. . . . . . . . . . . . . . . . . . . . . . 37Designated use. . . . . . . . . . . . . . . . . . . . . . . . . . . 4Device address . . . . . . . . . . . . . . . . . . . . . . . . . . 23Device description . . . . . . . . . . . . . . . . . . . . . . . 10Dimensions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33Documentation
Methods of ordering. . . . . . . . . . . . . . . . . . . . . 8
38
E
Electrical connection. . . . . . . . . . . . . . . . . . 22, 36Power supply. . . . . . . . . . . . . . . . . . . . . . 19, 36Sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21Signal output. . . . . . . . . . . . . . . . . . . . . . 19, 36Transmitter. . . . . . . . . . . . . . . . . . . . . . . . . . 17Electrical technician. . . . . . . . . . . . . . . . . . . . . . 17Electromagnetic compatibility . . . . . . . . . . . . . . 37Environment. . . . . . . . . . . . . . . . . . . . . . . . . . . 36
F
Factory setting
Access code. . . . . . . . . . . . . . . . . . . . . . . . 26–27
H
Housing
Closed. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Open. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Housing grounding . . . . . . . . . . . . . . . . . . . . . . 18Housing opening. . . . . . . . . . . . . . . . . . . . . . . . 17
I
Icons
Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Identification. . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Nameplate. . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Serial number . . . . . . . . . . . . . . . . . . . . . . . . . 6Incoming acceptance. . . . . . . . . . . . . . . . . . . . . . 4Input
Cable specification. . . . . . . . . . . . . . . . . . . . . 35Measured variables . . . . . . . . . . . . . . . . . . . . 35Measuring range . . . . . . . . . . . . . . . . . . . . . . 35Installation . . . . . . . . . . . . . . . . . . . . . . . . . . 4, 11Installation instructions
Dimensions. . . . . . . . . . . . . . . . . . . . . . . . . . 12Panel mounting. . . . . . . . . . . . . . . . . . . . . . . 15Wall or field mounting. . . . . . . . . . . . . . . . . . 14Installation kits . . . . . . . . . . . . . . . . . . . . . . . . . 34
L
Load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
M
Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . 29Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Liquiline M CM42
Maximum measured error . . . . . . . . . . . . . . . . . 36Measured value resolution . . . . . . . . . . . . . . . . . 36Measured variables. . . . . . . . . . . . . . . . . . . . . . . 35Measuring range. . . . . . . . . . . . . . . . . . . . . . . . . 35Measuring system. . . . . . . . . . . . . . . . . . . . . . . . 11Mechanical construction
Material. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37Weight. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37Mounting plate. . . . . . . . . . . . . . . . . . . . . . . . . . 13
N
Nameplate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
O
Opening the housing . . . . . . . . . . . . . . . . . . . . . 17Operation
Safety instructions . . . . . . . . . . . . . . . . . . . . . . 4Operational safety. . . . . . . . . . . . . . . . . . . . . . . . . 5Output
Load. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35Output distribution. . . . . . . . . . . . . . . . . . . . . 35Output signal. . . . . . . . . . . . . . . . . . . . . . . . . 35Signal on alarm. . . . . . . . . . . . . . . . . . . . . . . . 35
P
Performance characteristics
Maximum measured error . . . . . . . . . . . . . . . 36Measured value resolution . . . . . . . . . . . . . . . 36Post-installation
Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16Power supply . . . . . . . . . . . . . . . . . . . . . . . . 19, 36Product structure . . . . . . . . . . . . . . . . . . . . . . . . . 7
Q
Quick Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
R
Relative humidity. . . . . . . . . . . . . . . . . . . . . . . . 37Replacing modules . . . . . . . . . . . . . . . . . . . . . . . 30Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5, 33
S
Safety conventions and icons . . . . . . . . . . . . . . . . 5Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . 4Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . 8Sensor
Electrical connection . . . . . . . . . . . . . . . . . . . 21Serial number. . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Endress+Hauser
Signal output. . . . . . . . . . . . . . . . . . . . . . . . . 19, 36Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Storage temperature . . . . . . . . . . . . . . . . . . . . . . 37Symbols
Document. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
T
Technical data. . . . . . . . . . . . . . . . . . . . . . . . . . . 35Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35Mechanical construction. . . . . . . . . . . . . . . . . 37Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35Performance characteristics. . . . . . . . . . . . . . . 36Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . 30Types of calibration. . . . . . . . . . . . . . . . . . . . . . . 28
U
Use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4User administration. . . . . . . . . . . . . . . . . . . . . . . 26User groups. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27User roles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
W
Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Weather protection. . . . . . . . . . . . . . . . . . . . . . . 34Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
www.endress.com/worldwideBA381C/07/en/POD
Printed in Germany / FM+SGML 6.0