Operating Instructions

Liquiline M CM42

Two-wire transmitter for pH measurement with digital IS-FET sensors

Part 1: Commissioning

BA381C/07/en/PODValid as of:

Software version 10.04.00

Overview of documentation supplied

Operating Instructions

The Operating Instructions are split into two parts:•Commissioning

–Description of your device

–All steps that you have to carry out once only during initial commissioning•Operation

–Software description–Troubleshooting

–All steps that you have to carry out during the routine operation of your device, depending on your measuring task

!

Note!

This part of the Operating Instructions is the documentation on commissioning.

CD-ROM

Contents:

•Fundamentals of pH and conductivity measurement•Communication

•Overview of analytical measuring technology•Accessories•Safety

•Ordering information

In addition

•1 manufacturer's certificate•Adhesive labels:

–Replacement nameplate (landscape format)–Wiring adhesive label

Liquiline M CM42

Table of contents

1

Safety instructions . . . . . . . . . . 4

1.1 Designated use . . . . . . . . . . . . . . . . . . . . . . 41.2 Incoming acceptance, transport, storage . . . 41.3 Installation, commissioning and operation . 41.4 Operational safety . . . . . . . . . . . . . . . . . . . 51.5 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51.6

Notes on safety conventions and icons . . . . 5

2

Identification . . . . . . . . . . . . . . 6

2.1 Device designation . . . . . . . . . . . . . . . . . . . 62.2 Scope of delivery . . . . . . . . . . . . . . . . . . . . 82.3 Reordering documentation . . . . . . . . . . . . . 82.4

Certificates and approvals . . . . . . . . . . . . . . 9

3 Device description . . . . . . . . . 103.1 Housing closed . . . . . . . . . . . . . . . . . . . . . 103.2 Housing open . . . . . . . . . . . . . . . . . . . . . . 10

4

Installation. . . . . . . . . . . . . . . 11

4.1 Measuring system . . . . . . . . . . . . . . . . . . 114.2 Installation conditions . . . . . . . . . . . . . . . 124.3 Installation instructions . . . . . . . . . . . . . . 144.4

Post-installation check . . . . . . . . . . . . . . . 16

5 Wiring . . . . . . . . . . . . . . . . . . 17

5.1 Electrical connection . . . . . . . . . . . . . . . . 175.2 Post-connection check . . . . . . . . . . . . . . . 22

6

Commissioning. . . . . . . . . . . . 23

6.1 Function check . . . . . . . . . . . . . . . . . . . . 236.2 Configuring the device address . . . . . . . . . 236.3 Quick Setup (local operation) . . . . . . . . . . 236.4 User administration . . . . . . . . . . . . . . . . . 266.5

Calibration . . . . . . . . . . . . . . . . . . . . . . . . 27

7 Maintenance. . . . . . . . . . . . . . 298

Troubleshooting - hardware. . 30

8.1 Replacing modules . . . . . . . . . . . . . . . . . . 308.2 Spare parts . . . . . . . . . . . . . . . . . . . . . . . . 318.3 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . 338.4

Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . 33

Endress+Hauser

9

Accessories. . . . . . . . . . . . . . .34

9.1 Installation kits . . . . . . . . . . . . . . . . . . . . . 349.2 Weather protection . . . . . . . . . . . . . . . . . . 349.3 Communication . . . . . . . . . . . . . . . . . . . . 349.4

Data management . . . . . . . . . . . . . . . . . . . 34

10 Technical data. . . . . . . . . . . . .35

10.1 Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3510.2 Output . . . . . . . . . . . . . . . . . . . . . . . . . . . 3510.3 Electrical connection . . . . . . . . . . . . . . . . . 3610.4 Performance characteristics . . . . . . . . . . . . 3610.5 Environment . . . . . . . . . . . . . . . . . . . . . . 3610.6

Mechanical construction . . . . . . . . . . . . . . 37

Index. . . . . . . . . . . . . . . . . . . .38

Safety instructionsLiquiline M CM42

1

1.1

Safety instructions

Designated use

Liquiline CM42 is a two-wire transmitter for liquid analysis in all areas of process engineering:•Chemical processes•Pharmaceutical industry•Food technology

If the device is used for any purpose other than that described, this poses a threat to the safety of people and the entire measuring system and is thus not permitted.

The manufacturer does not accept liability for damage caused by improper or non-designated use.

1.2Incoming acceptance, transport, storage

•Make sure the packaging is not damaged!

Inform your supplier of any damage to the packaging.

Please keep the damaged packaging until any issues have been resolved.•Make sure the contents are not damaged!

Inform your supplier of any damage to the contents.

Please keep the damaged goods until any issues have been resolved.

•Check the scope of delivery against the delivery papers and your order to ensure it is complete and nothing is missing.

•Pack the product in such a way as to protect it reliably against impact and moisture for storage and transportation. Optimum protection is provided by the original packaging materials. In addition, the permitted ambient conditions must be observed (see Technical data).•If you have any queries, please contact your supplier or local sales center.

1.3Installation, commissioning and operation

Observe the following points:

•Installation, commissioning, operation and maintenance of the measuring system must only be carried out by trained technical personnel.

The technical personnel must be authorized to perform the tasks by the owner-operator.•The electrical connection may only be established by an electrical technician.

•The technical personnel must have read and understood these Operating Instructions and must follow the instructions they contain.

•Prior to commissioning the entire measuring point, check that all connections are correct. Make sure that electric cables and hose connections are not damaged.

•Do not commission damaged products. Protect them against unintentional startup. Label and identify the damaged product as defective.

•Faults at the measuring point may only be rectified by authorized and properly trained personnel.

4Endress+Hauser

Liquiline M CM42Safety instructions

•If the faults cannot be eliminated, take the products out of service and protect them against unintentional startup.

•Repairs not described in these Operating Instructions may only be carried out directly at the manufacturer's or by the Service Organization.

1.4Operational safety

The transmitter is designed to meet state-of-the-art safety requirements, has been tested and left the factory in a condition in which it is safe to operate.

The applicable regulations and European standards have been taken into account.As the user, you are responsible for observing the following safety regulations:•Installation guidelines

•Local standards and regulations

1.5Return

If returning the transmitter, please send it cleaned to your sales center.Use the original packaging when returning the device.

1.6Notes on safety conventions and icons

1.6.1Warnings

Warning!

This symbol alerts you to dangers.

If the warning is not observed, this can result in serious injury to people or damage to property.

Caution!

This symbol alerts you to possible malfunctions resulting from incorrect operation. If not observed, this could result in damage to property.Note!

This symbol indicates important items of information.

#\"!

1.6.2Document symbols

È ä 1This symbol stands for a cross-reference to a certain page (e.g. Page 1).È å 2This symbol stands for a cross-reference to a certain graphic (e.g. Fig. 2).

Endress+Hauser5

IdentificationLiquiline M CM42

2

2.1

Identification

Device designation

2.1.1Nameplate

Nameplates can be found:

•On the outside of the housing (lasered)

•On the packaging (adhesive label, portrait format)

•On the adhesive label sheet provided (landscape format, for your own use)Compare the data on the nameplate with your order.

The nameplate provides you with the following information on your device:1.Order code2.Serial number3.Device version4.Measuring range(s)5.Input, output values

6.Approvals (depending on the version ordered)7.Protection class

8.Ambient temperature range9.Approval symbols

a0003281Fig. 1: Nameplate information

10.Safety notices, warnings

2.1.2Serial number and order code

The order code and serial number of your device can be found in the following locations:•On the nameplate

•On the front page of the Operating Instructions•In the delivery papers

!

Note!

•The documentation only contains information that was valid at the time of delivery.•If you have any queries, please contact your local sales center.

6Endress+Hauser

Liquiline M CM42Identification

2.1.3Product structure

Sensor inputANomoduleCConductivity, conductive measurement, analog sensorIConductivity, inductive measurement, analog sensorKDigital sensor: conductivity, conductive measurementMDigital sensor: pH/ORP measurement, glass sensorNDigital sensor: pH measurement, ISFET sensorODigital sensor: amperometric oxygen measurementPpH (glass/ISFET) or ORP, analog sensor

ApprovalANon-hazardous area, CSA GPGAtex II (1)2 G EEx ia/ib IIC T6 / II 3 DHAtex II 3G EEx nA II T6NNEPSIExnAIIT6PFM IS NI Cl. I, Div. 1&2, Groups A-D

SCSA IS NI Cl. I, II, III, Div. 1&2, Groups A-G

Certificate ANot selectedBTest certificate to EN10204, 3.1CTest certificate to EN10204, 3.1, factory calibration certificate

Output

® 01 current output 4 to 20 mA, HART

12 current outputs 4 to 20 mA, 1x HART®2PROFIBUSPA3FOUNDATION Fieldbus

Housing

0Plastic housing1Stainless steel housing 1.4404 (AISI 316L)

Cable entry0M20 x 1.51NPT½\"2G½

Software EAStandard versionEBAdvanced versionECAdvanced Plus version

Device languageDDe/En(Docs,German)EEn/De (Docs, English)FFr/En(Docs,French)SEs/En (Docs, Spanish) ZZh/En (Docs, Chinese)

Documentation0Installation + operation

Additional fittings0Basic version

Complete order code

CM42-!

Note!

Please ask your sales office for other device versions, e.g. other software or documentation languages.

Endress+Hauser7

IdentificationLiquiline M CM42

2.2

••••••

Scope of delivery

1 transmitter in the version ordered

1 wall securing unit incl. 4 flat head screws

1 sheet of adhesive labels (nameplates, terminal connection diagrams)1 Operating Instructions Part 1, BA381C \"Commissioning\"1 Operating Instructions Part 2, BA382C \"Operation\"1 CD-ROM with additional documentation

If you have any queries, please contact your supplier or local sales center.

2.3Reordering documentation

What to order?

•Reordering lost documentation or documentation of which several copies are required:You must quote the serial number of your device.

You receive the same documentation as received with the delivery.

•First time ordering appropriate documentation for devices which you have assembled yourself by installing a module:

You have to indicate the order code the device corresponds to after installing the module. Specify a serial number also.

You receive the documentation to suit your newly modified device.How to order?

•By Internet (www.endress.com/liquiline-documentation →å2):You receive PDF files of both parts of the Operating Instructions.•Through your sales center:

Indicate you want the documentation as hard copy.

a0006062

Fig. 2: www.endress.com/liquiline-documentation

8Endress+Hauser

Liquiline M CM42Identification

2.4Certificates and approvals

With this declaration, the manufacturer guarantees that the product conforms to the regulations of European Directive 89/336/EEC. This is proven by observing the standards listed in the declaration of conformity.

Endress+Hauser9

Device descriptionLiquiline M CM42

3

3.1

Device description

Housing closed

1345-8

Wall securing plateAlarm LEDDisplayNavigatorSoft keys,

assignment depends on menu

122

348Fig. 3: CM42 with closed display cover

765a00010673.2Housing open

a0001069Fig. 4: 123With open display cover (without wiring)DAT memory moduleCPU moduleSensor module

456DIN rail

Threaded joints

CDI socket (service interface)

10Endress+Hauser

Liquiline M CM42Installation

4

4.1

Installation

Measuring system

A complete measuring system consists of:•Liquiline M CM42

•An assembly, e.g. fixed installation assembly CPA442•A digital sensor, e.g. Tophit CPS471-D•A measuring cable CYK10

a0001399Fig. 5: Example of a measuring system1234

Liquiline M CM42

Measuring cable CYK10Digital sensor

Fixed installation assembly CPA442

Endress+Hauser11

InstallationLiquiline M CM42

4.2Installation conditions

4.2.1Dimensions

a0001074Fig. 6: Dimensions

12Endress+Hauser

Liquiline M CM42Installation

4.2.2Mounting plate

a0001075Fig. 7: Mounting plate

4.2.3Weather protection cover

!

Note!

When installing outside, always use the weather protection cover (see 'Accessories', CD-ROM) to protect the transmitter against rain and snow.

107(4.21)90°253(9.96)224(8.82)98(3.86)151(5.94)Fig. 8: Weather protection cover

98(3.86)mm(inch)a0001671Endress+Hauser

230(9.01)100°13

InstallationLiquiline M CM42

4.3Installation instructions

4.3.1Wall or field mounting

There are many ways of mounting the device:•Wall mounting

•Mounting on a vertical pipe or post (round or square)

•Mounting on railing or on a horizontal pipe (round or square)

!

Note!

You require the post mount to mount on a pipe or post. This is an accessory and is not included in the scope of delivery (see \"Accessories\").

121244Fig. 9: Mounting on a vertical pipe/post12

CM42 (as delivered, incl. a mounting plate)

Securing screws (part of the post mount, see Accessories)

3a00030923a0002162Fig. 10: Side view34

Mounting plate (post mount)Pipe or post (round/square)

14Endress+Hauser

Liquiline M CM42Installation

121244Fig. 11: Mounting on a horizontal pipe/railing12

CM42 (as delivered)

Securing screws (post mount)

3a00030933a0002164Fig. 12: Side view34

Mounting plate (post mount)Pipe or railing

a0002166Fig. 13: Wall mounting (optional weather protection cover)

4.3.2Panel mounting

!

Note!

•For panel mounting, you need the installation kit consisting of tensioning screws and a front seal. This is an accessory (see \"Accessories\") and is not included in the scope of delivery.•If installing the units above one another, you must observe a minimum distance for the cable glands of the upper device.

•If installing the units beside one another, you must observe a minimum distance for opening the front of the housing.

•If arranging in a square, you must take into account the lengths of the mounting plates at the rear of the device or the cable glands for minimum spacing distance.

Endress+Hauser15

InstallationLiquiline M CM42

a0002402Fig. 14: Panel mounting: left side view, right front view

4.4Post-installation check

•After installation, check the transmitter for damage.

•Check whether the transmitter is protected against moisture and direct sunlight.

16Endress+Hauser

Liquiline M CM42Wiring

5

5.1

Wiring

Electrical connection

#

Warning!

•The electrical connection may only be established by an electrical technician.

•The electrical technician must have read and understood these Operating Instructions and must follow the instructions they contain.

•Prior to beginning any wiring work, make sure voltage is not applied to any of the cables.5.1.1Opening the housing1.

Loosen the four screws on the front with a Phillips screwdriver:

2.Open the housing:

\"

Caution!

Do not use any sharp or pointed objects, such as a screwdriver, knife etc., to open the housing. Otherwise, you might damage the housing seal, scratch the housing etc.

Endress+Hauser17

WiringLiquiline M CM42

5.1.2Housing grounding

\"

Caution!

You must connect the \"fixing plate\" to the foundation ground with a separate ≥2.5mm2 (i14 AWG) functional ground.

a0003617Fig. 15: Grounding the housing

12

Fixing plate

≥2.5mm2 (i14 AWG) functional ground

5.1.3Cable grounding in housing

!

Note!

•If possible, only use terminated genuine cables.

•Ground all the cables in the housing of the transmitter as shown in the following diagrams (sample cable, does not necessarily correspond to the original cable).123a0002679a0002680a0002681Fig. 16: Terminated cable (example)123

Outer shield (exposed)Cable cores with ferrulesCable sheath (insulation)

Fig. 17: Inserting the cable4

Grounding clip

Fig. 18: Grounding the cable

(functional ground)

18Endress+Hauser

Liquiline M CM42Wiring

5.1.4Cable terminals

1.Insert a suitable screwdriver into the opening

of the terminal spring (square opening) until the stop.

a0003602Fig. 19: Opening the terminal

2.Insert the terminated cable end into the

terminal opening (round opening).

a0003603Fig. 20: Inserting the cable

3.Remove the screwdriver. Make sure the cable

is securely positioned in the terminal.

a0003604Fig. 21: Releasing the terminal spring

5.1.5Signal output / power supplyPermitted supply voltage

Current output 4 to 20 mA without HART®:Current output 4 to 20 mA with HART®:

1)

12.5 to 30 V DC12.5 to 30 V DC1)

Permitted residual ripple to IEC 60381-1:1985 Par. 3.3: < 100 mVpp in frequency range 50 Hz to 3 kHz

Connection in transmitter

!

Note!

Connect the transmitter with two-wire cabling shielded on both sides.

Endress+Hauser19

WiringLiquiline M CM42

a0002365Fig. 22: View in device (CPU module)

a0003100Fig. 23: Wiring diagram

Connection via HART interface

12RS232³250W+-1331343a0004595Fig. 24: Connection: HART without PLC123

Power unit

HART modem Commubox FXA191HART handheld terminal

20Endress+Hauser

Liquiline M CM42Wiring

1+-1331342RS2323a0004597Fig. 25: Connection: HART with PLC123

Additional switching units or PLC with passive inputHART modem Commubox FXA191HART handheld terminal

5.1.6Sensor connection

Explanation of abbreviations in the following diagrams:AbbreviationU+U–

Com ACom B

Meaning

Signals of the digital sensor

!

Note!

•Cable colors indicated as per IEC 757 (see CD–ROM).

•You must connect shielded connections and terminals with functional ground ()) (there is no protective ground (*) for plastic housings).

Endress+Hauser21

WiringLiquiline M CM42

a0006128a0001078Fig. 26: View in device (sensor module)Fig. 27: Wiring diagram

5.2

1.2.3.4.5.6.7.

Post-connection check

Are the transmitter and cables free from damage on the outside?Are the mounted cables strain relieved?Are the cables run without loops and cross-overs?

Are the signal lines correctly connected in accordance with the wiring diagram?Have you positioned unused connection wires on the \"fixing plate\"?Are all the connection wires securely positioned in the cable terminals?Are all the cable entries installed, tightened and sealed?

Device state and specificationsElectrical connection

22Endress+Hauser

Liquiline M CM42Commissioning

#

Warning!

Only put the transmitter into operation if you can answer yes to all these questions. Otherwise, safe operation cannot be guaranteed. In such instances, the manufacturer does not accept any liability for the device.

6

6.1

Commissioning

Function check

#

Warning!

•Check that all connections have been established correctly.

•Make sure that the supply voltage matches the voltage indicated on the nameplate!

6.2Configuring the device address

Factory setting: 0.

You can change the device address to integrate several HART devices in a network (Multidrop mode).

Use the handheld terminal (DXR375) to change the device address.

6.3

1.2.3.

Quick Setup (local operation)

Connect the supply voltage.

Wait for the initialization to be complete.

Completely work through the \"Quick Setup\" submenu in \"PARAM\".

6.3.1Menu structure

Function name of display (local operation)PARAMQuick Setup

LanguageDate formatSet dateTime formatSet timeTemperature unitBuffer manufacturerTemp. compensation

Endress+Hauser23

CommissioningLiquiline M CM42

Function name of display (local operation)Medium temperature(Temp. compensation=\"Man. comp.\")Current output 1

Output sourceLow value (4mA)Upper value (20mA)Current output 2

Output sourceLow value (4mA)Upper value (20mA)

6.3.2Configuration options

The abbreviation \"AC\" is used in the table header. AC stands for \"Access Controls\write authorization.

The table only indicates the rights of the \"Maintenance\" user group as the \"Specialist\" group has administrator status and thus always has full rights (--> \"Commissioning/User administration\").•R = Read only, i.e. the maintenance engineer can only read the data

•R/W = Read/Write, i.e. as a maintenance engineer, you can also change values.

!

Note!

The access controls do not take effect until password protection has been activated!

FunctionLanguage

Options

Options•English

•Language orderedFactory settingLanguage ordered

Date format

Options

•DD.MM.YYYY•MM.DD.YYYYFactory settingDD.MM.YYYY

Set dateTime format

Depends on the formatDD.MM.YYYYOptions

•hhmmss (24 h)•hhmmss (am / pm)Factory settinghhmmss (24 h)

Set time

Depends on the formathh:mm:ss

R/WR/WR/W

24-hour display or 12-hour displayEditing mode:

hh (hour): 0 to 23 / 0 am to 12 pmmm (minutes): 0 to 59ss (seconds): 0 to 59

R/WACR/W

Info

Note! !\"Language ordered\" refers to the language that you selected for your device using the order code (\"device language\").

If you select the other language in question, all the other settings remain intact.Editing mode:DD (day): 1 to 31MM (month): 1 to 12

YYYY (year): 2005 to 2100

24Endress+Hauser

Liquiline M CM42Commissioning

FunctionTemperature unit

OptionsOptions• °C• °F

Factory setting °C

ACR/W

Info

Buffer manufacturer

Options

•E+H(NIST)•Ingold/Mettler•DIN 19267•DIN 19266•Merck/Riedel Factory settingE+H (NIST)

R/W

Temp. compensation

Options•Off

•Autocomp.(ATC)•Man.comp.Factory settingAuto comp. (ATC)

R/W

Medium temperature

–50 to + 250 °C

(–58 to + 482 °F)Factory setting25 °C (77 °F)

R/W

Only for temp. compensation = \"Man. comp.\"

Current output 1Output source

Options

•Mainvalue

Factory setting Main value

–2 to

(20 mA value) - 1 pHFactory setting2 pH

Upper value (20mA)

(4 mA value) + 1 pH to16 pHFactory setting12 pH

Current output 2Output source

Options

•Mainvalue•TemperatureFactory settingTemperature

RR/WR

Main value

Low value (4mA)R/W

Specify the measured value that should correspond to the 4mA or 20mA value.The smallest possible spread between4mA and 20mA value is:1 pH

Temperature

Endress+Hauser25

CommissioningLiquiline M CM42

Function

Low value (4mA)

Options

–50 to

(20 mA value) - 5 °CFactory setting-20 °C

ACR/W

Info

The smallest possible spread between 4mA and 20mA value is 0.1 °C.

Upper value (20mA)

(4 mA value) + 5 °C to250 °CFactory setting+100 °C

R/W

6.4User administration

To protect the device against unwanted or unauthorized access, you can assign an access code to the menu.

!

Note!

The factory setting for the access code is an empty field (= no entry). The prompt to enter a password is deactivated. Setting is possible for everyone.Assigning/changing access codes:1.2.

Go to PARAM menu, \"General settings\"/\"User administration\".Select the type of protection:

–No password protection (= factory setting).–Password protection per code entry.

If you select \"Enter code\roles.

3.

6.4.1User roles and access authorization

If the password protection has been defined and activated, you have to log in with the appropriate user role before you can make changes to the settings in the menu.

26Endress+Hauser

Liquiline M CM42Commissioning

Here, you will be prompted by the device to enter an access code as soon as you want to configure a value:

User roles and access authorization (factory setting):

Maintenance

•Read and, sometimes, write access in the PARAM and DIAG menus

•Read and write access in the calibration menu

Specialist

Admin status:

•Read and write access in all menus

•Change the access authorization for other user groups

!

Note!

If you are logged in as the \"Specialist\" (= Administrator), you can change the access authorization of all other user groups:1.2.

Select a software function for which you want to assign authorization to, or take authorization from, another user group.

Press and hold down the navigator button for at least 3 seconds.

The display shows all the user groups and the current access authorization to the function selected.

Use the navigator to change the authorization in question and confirm with \"OK\".Make a note of the changes in your operating logbook or similar.

3.4.

6.5Calibration

Calibration is necessary:•After periods of standstill

•At sensible, process-dependent intervals1)

1) Depending on the process conditions, the intervals can range from several times daily to once quarterly.

Endress+Hauser27

CommissioningLiquiline M CM42

6.5.1Types of calibration

•Two-point calibration

–With automatic buffer recognition:

You select the buffer type (e.g. DIN 19266). The device automatically recognizes the buffer during calibration.–With fixed buffer:

You select two buffer solutions from the menu.•Single-point calibration –You enter an offset.•Numeric input:

–You enter the slope, zeropoint and temperature.•Grab sample calibration

–You take a sample of the medium. The device determines a measured value while you analyze the sample in the laboratory.

Then you enter the laboratory measured value. The device calibrates on the basis of the laboratory value.6.5.2Calibrate

Proceed as follows to calibrate the sensor:1.2.3.4.5.6.

Remove the sensor from the process.Clean the sensor.

Press the soft key for \"CAL\".Follow the instructions in the menu.

Finish calibrating by switching back to the measuring mode.Install the sensor back into the process.

Your measuring point is now ready for operation.

!

Note!

•If calibration is aborted using ESC, or if the calibration is faulty, the system continues to use the original calibration data. A calibration error is shown as plain text on the display. •Any offset set is automatically deleted after accepting the calibration.

28Endress+Hauser

Liquiline M CM427Maintenance

Clean the front of the housing with usual commercial cleaning agents.The front is resistant to the following in accordance with DIN 42 115:•Alcohol (briefly)

•Diluted acids (max. 2% HCl)•Diluted alkalis (max. 3% NaOH)•Soap-based household cleaner

\"

Caution!

Never use any of the following for cleaning purposes:•Concentrated mineral acids or alkalis•Benzyl alcohol•Methylene chloride•High-pressure steam

Endress+HauserMaintenance

29

Troubleshooting - hardwareLiquiline M CM42

8

8.1

Troubleshooting - hardware

Replacing modules

Exchange modules when replacing defective modules or in the event of changes to the configuration of your device.

112a0004575a0004577Fig. 28: Removing module1

Removal aids

Fig. 29: Housing guides12

Guides on housing wallGuidesinDINrail 1a0004578a0004579Fig. 30: Module, left side (cable connections = front)1

Guides (e.g. CPU module to fit into the guides in

housing)

Fig. 31: Module, right side (cable connection = front)12

Guides (e.g. input module to fit into the right-hand guides of the CPU module)Removal aids

1.2.3.4.5.6.7.

Disconnect Liquiline from the power supply and open the housing.Remove the cable connections from the module you want to replace.Pull out the two removal aids on the module until the stop (→å28).The module can now be easily removed from the DIN rail.

Slide the new module into the guides (→å29, →å30, →å31).

Push the two removal aids on the module in the direction of the DIN rail until the stop. This locks the module onto the DIN rail.

Connect the cables in accordance with the wiring diagram (see \"Wiring\").

Close the housing, connect the sensor and check that the entire measuring system is working correctly.

30Endress+Hauser

Liquiline M CM42Troubleshooting - hardware

8.2Spare parts

a0004474Fig. 32: Exploded view

Please refer to the following table for item names and order numbers for spare parts.

Endress+Hauser31

Troubleshooting - hardwareLiquiline M CM42

Item1

Kit CM42

Housing base, plastic, M20 1)•Base

•Plate M20 (item 15)•Fixing plate cpl. SS

•3 threaded joints M20x1.5 (item 14)

CPU module, curr. outp. + HART, non-Ex

•FMIH1 insertion module (FC2W1+FBIH1), non-Ex•Terminal strip (item 3)

Terminal set, CPU module, curr. outp. + HARTRibbon cable for plastic housing

•1 cable, conf. 40xAWG28 EX, 90 mm (3.5\") crimpHinge for plastic housing•Hinge cpl. CM42 KS

•2 special screws K40x10 Torx self-form.DATchipwith software

•FMDC1 FTC electronics BG EX

Top housing section, plastic, with display

•Display, keys, navigator, screws and cover plate•Hinge(item5)

•Ribbon cable (item 4)

Inp. module conductivity, conductive measurement, non-Ex•FSLC1, insertion module, cond. conductivity, non-Ex

•Terminal strip (item 9)

Order number71001361

251518002

345

515174815151750351517500

67

5151750951517462

851518005

8

Inp. module conductivity, inductive measurement, non-Ex •FSLI1, insertion module, ind. conductivity, non-Ex

•Terminal strip (item 9)

Inp. module terminal set, conductivity, conductive measurementInp. module terminal set, conductivity, inductive measurementInp. module digital/Memosens, non-Ex

•FSDG1, insertion module, 1 channel, 1 SWU, non-Ex•Terminal strip (item 11)

Inp. module terminal set, digital/MemosensInp. module pH/ORP/temperature, non-Ex•FSPH1, insertion module, pH/mV, non-Ex•Terminal strip (item 13)

Inp. module terminal set, pH/ORP/temperatureThreaded joints M20•1 set of threaded joints•Thread base plate

Terminal set, 5 pcs. each for all modules, 7 types

51518006

9910

515174895151749051518007

1112

5151749151518004

1314

5151748751517504

No Fig.51517498

32Endress+Hauser

Liquiline M CM42Troubleshooting - hardware

ItemKit CM42

Order numberNo Fig.

10 set cable clamps+screws

51517499

•10 EMC cable clamps D6 (0.24\")•10 EMC cable clamps D4 (0.16\")•10 cheese head screws M4x8 A2No Fig.Connection jack for external Historom/CDI

51517507

1)

When ordering, you must specify the serial number of the device for which you are ordering the spare part.

8.3Return

If returning the transmitter, please send it cleaned to your sales center.Use the original packaging when returning the device.

8.4Disposal

This product contains electronic components. For this reason, it must be disposed of as electronic waste.

Please observe local regulations.

Endress+Hauser33

AccessoriesLiquiline M CM42

9Accessories

!

Note!

The CD-ROM supplied provides you with all the relevant information on an extensive range of accessories, including descriptions, graphics and ordering information.For example, it contains information on:•Sensors•Assemblies

•Quality buffers from Endress+Hauser

9.1Installation kits

Post mount

•1 Mounting plate•2 Securing screws•Order No. 51518263

Panel installation kit for panel cutout 138x138 mm (5.43x5.43 inch)•1 Panel installation seal

•2 Tensioning screws M6x150 mm•4 Hexagonal nuts M6, DIN934 A2•4 Spring washers, A2 DIN127, form B6•4 Washers A6.4, DIN125 A2•Order No. 51518173

9.2Weather protection

Weather protection cover •Order No. 51517382

9.3Communication

HART handheld terminal DXR375

•For communicating with a HART-compatible device via a 4 to 20 mA line•Order No. DXR 375

HART modem Commubox FXA291

•Interface module between HART and serial PC interface•Technical Information TI405C/07•Order No. 51516983

9.4Data management

Memobase CYZ41D

•Software for central data and sensor management•for Windows 2000 or XP

•Ordering as per product structure

34Endress+Hauser

Liquiline M CM42Technical data

10Technical data

10.1Input

10.1.1Measured variables•pH value•Temperature

10.1.2Measuring rangepH valueTemperature

0 to 14

–25 to +150 °C (–10 to 300 °F)

10.1.3Cable specificationWith Memosens (CYK10)

Max. cable length 100 m (330 ft)

10.2Output

10.2.1Output signal

2x 4 to 20 mA, potentially isolated against sensor circuit0.8 to 1.2 mA peak to peak2)10.2.2Signal on alarm≥21.5 mA10.2.3Load

Max. load with supply voltage of 24V: 500 ΩMax. load with supply voltage of 30 V: 750 Ω10.2.4Output distributionpH valueTemperature

Configurable, ΔpH > 0.5Adjustable, at least 2 °C (2 °F)

2) Only for HART

Endress+Hauser35

Technical dataLiquiline M CM42

10.3Electrical connection

10.3.1Supply voltage and signal voltageCurrent output 4 to 20 mA without HART®:Current output 4 to 20 mA with HART®:

1)

12.5 to 30 V DC12.5 to 30 V DC1)

Permitted residual ripple to IEC 60381-1:1985 Par. 3.3: < 100 mVpp in frequency range 50 Hz to 3 kHz

10.3.2Cable cross-section

Max. cable cross-section: 2.5 mm2 (i14 AWG), GND 4 mm2 (i12 AWG)

10.4Performance characteristics

10.4.1Measured value resolutionpH

Leakage current, ISFET sensorsTemperature

0.01 pH100 nA0.1 °C (0.1 °F)

10.4.2Maximum measured error3)Current outputs, additionalpH

Leakage current, ISFET sensorsTemperature

20 μA0.02 pH100 nA1 K

10.5Environment

10.5.1Ambient temperature range–30 to 70 °C (–20 to 160 °F)

3) In accordance with IEC 746-1 for nominal operating conditions

36Endress+Hauser

Liquiline M CM42Technical data

10.5.2Ambient temperature limits–30 to +80 °C (–20 to 175 °F)10.5.3Storage temperature–40 to 80 °C (–40 to 175 °F)

10.5.4Electromagnetic compatibility

Interference emission and interference immunity to EN 61326: 200410.5.5Degree of protectionIP67 (iNEMA 4X)10.5.6Relative humidity10 to 95%, not condensing

10.6Mechanical construction

10.6.1Weight1.5 kg (3.3 lbs)10.6.2MaterialHousing

Housing seals

PolycarbonateEPDM

Endress+Hauser37

Liquiline M CM42

Index

A

Access authorization

Change. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26User roles. . . . . . . . . . . . . . . . . . . . . . . . . . . . 26Access code

Factory setting . . . . . . . . . . . . . . . . . . . . . . . . 26Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34Communication . . . . . . . . . . . . . . . . . . . . . . . 34Installation kits. . . . . . . . . . . . . . . . . . . . . . . . 34Ambient temperature limits. . . . . . . . . . . . . . . . . 37Ambient temperature range. . . . . . . . . . . . . . . . . 36Approvals. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

B

Bus address. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23

C

Cable cross-section. . . . . . . . . . . . . . . . . . . . . . . 36Cable grounding. . . . . . . . . . . . . . . . . . . . . . . . . 18Cable specification . . . . . . . . . . . . . . . . . . . . . . . 35Cable terminals. . . . . . . . . . . . . . . . . . . . . . . . . . 19Calibrate. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27Check

Electrical connection. . . . . . . . . . . . . . . . . . . . 22Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23Post-installation . . . . . . . . . . . . . . . . . . . . . . . 16Commissioning. . . . . . . . . . . . . . . . . . . . . . . . 4, 23Communication . . . . . . . . . . . . . . . . . . . . . . . . . 34Configuring

Device address . . . . . . . . . . . . . . . . . . . . . . . . 23Connecting ground. . . . . . . . . . . . . . . . . . . . . . . 18Connection

Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22Connection data. . . . . . . . . . . . . . . . . . . . . . . . . 36

D

Degree of protection. . . . . . . . . . . . . . . . . . . . . . 37Designated use. . . . . . . . . . . . . . . . . . . . . . . . . . . 4Device address . . . . . . . . . . . . . . . . . . . . . . . . . . 23Device description . . . . . . . . . . . . . . . . . . . . . . . 10Dimensions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33Documentation

Methods of ordering. . . . . . . . . . . . . . . . . . . . . 8

38

E

Electrical connection. . . . . . . . . . . . . . . . . . 22, 36Power supply. . . . . . . . . . . . . . . . . . . . . . 19, 36Sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21Signal output. . . . . . . . . . . . . . . . . . . . . . 19, 36Transmitter. . . . . . . . . . . . . . . . . . . . . . . . . . 17Electrical technician. . . . . . . . . . . . . . . . . . . . . . 17Electromagnetic compatibility . . . . . . . . . . . . . . 37Environment. . . . . . . . . . . . . . . . . . . . . . . . . . . 36

F

Factory setting

Access code. . . . . . . . . . . . . . . . . . . . . . . . 26–27

H

Housing

Closed. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Open. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Housing grounding . . . . . . . . . . . . . . . . . . . . . . 18Housing opening. . . . . . . . . . . . . . . . . . . . . . . . 17

I

Icons

Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Identification. . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Nameplate. . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Serial number . . . . . . . . . . . . . . . . . . . . . . . . . 6Incoming acceptance. . . . . . . . . . . . . . . . . . . . . . 4Input

Cable specification. . . . . . . . . . . . . . . . . . . . . 35Measured variables . . . . . . . . . . . . . . . . . . . . 35Measuring range . . . . . . . . . . . . . . . . . . . . . . 35Installation . . . . . . . . . . . . . . . . . . . . . . . . . . 4, 11Installation instructions

Dimensions. . . . . . . . . . . . . . . . . . . . . . . . . . 12Panel mounting. . . . . . . . . . . . . . . . . . . . . . . 15Wall or field mounting. . . . . . . . . . . . . . . . . . 14Installation kits . . . . . . . . . . . . . . . . . . . . . . . . . 34

L

Load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35

M

Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . 29Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37

Liquiline M CM42

Maximum measured error . . . . . . . . . . . . . . . . . 36Measured value resolution . . . . . . . . . . . . . . . . . 36Measured variables. . . . . . . . . . . . . . . . . . . . . . . 35Measuring range. . . . . . . . . . . . . . . . . . . . . . . . . 35Measuring system. . . . . . . . . . . . . . . . . . . . . . . . 11Mechanical construction

Material. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37Weight. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37Mounting plate. . . . . . . . . . . . . . . . . . . . . . . . . . 13

N

Nameplate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

O

Opening the housing . . . . . . . . . . . . . . . . . . . . . 17Operation

Safety instructions . . . . . . . . . . . . . . . . . . . . . . 4Operational safety. . . . . . . . . . . . . . . . . . . . . . . . . 5Output

Load. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35Output distribution. . . . . . . . . . . . . . . . . . . . . 35Output signal. . . . . . . . . . . . . . . . . . . . . . . . . 35Signal on alarm. . . . . . . . . . . . . . . . . . . . . . . . 35

P

Performance characteristics

Maximum measured error . . . . . . . . . . . . . . . 36Measured value resolution . . . . . . . . . . . . . . . 36Post-installation

Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16Power supply . . . . . . . . . . . . . . . . . . . . . . . . 19, 36Product structure . . . . . . . . . . . . . . . . . . . . . . . . . 7

Q

Quick Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . 23

R

Relative humidity. . . . . . . . . . . . . . . . . . . . . . . . 37Replacing modules . . . . . . . . . . . . . . . . . . . . . . . 30Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5, 33

S

Safety conventions and icons . . . . . . . . . . . . . . . . 5Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . 4Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . 8Sensor

Electrical connection . . . . . . . . . . . . . . . . . . . 21Serial number. . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

Endress+Hauser

Signal output. . . . . . . . . . . . . . . . . . . . . . . . . 19, 36Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Storage temperature . . . . . . . . . . . . . . . . . . . . . . 37Symbols

Document. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

T

Technical data. . . . . . . . . . . . . . . . . . . . . . . . . . . 35Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35Mechanical construction. . . . . . . . . . . . . . . . . 37Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35Performance characteristics. . . . . . . . . . . . . . . 36Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . 30Types of calibration. . . . . . . . . . . . . . . . . . . . . . . 28

U

Use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4User administration. . . . . . . . . . . . . . . . . . . . . . . 26User groups. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27User roles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

W

Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Weather protection. . . . . . . . . . . . . . . . . . . . . . . 34Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

www.endress.com/worldwideBA381C/07/en/POD

Printed in Germany / FM+SGML 6.0